    Microfluidics 6: PDMS Chip, Optical Quality Control and Flow Leakage Test

    C. Yunus Sahan, Serhat Sevli

    Published: 2021-08-23 DOI: 10.17504/protocols.io.bxkvpkw6
    AI 解读
    浏览 1818

    Abstract

    Fabrication of microfluidic chips requires high quality and high resolution manufacturing techniques. After fabrication, the quality inspection is crucial and can be performed optically for visual parameters and flow application. If the sizes are well, chips are clean and intake, and flow is laminar without leakage, then everything is OK. This protocol describes the methods for testing of the chips fabricated in our laboratory.

    Before start

    This protocol is derived from open scientific literature and can be modified by user freely.

    Steps

    Chip Quality Control

    1.

    Optical Quality Control of the microfluidic chips are performed by using digital microscope.

    The visual parameters checked for acceptance of the chip are as listed below.

    • Defects and scratchs
    • Strains and voids
    • Dirt in features
    • Air bubbles in features
    • Ports and Bonding, out of alignment
    • Not bond features
    • Incorrect sizes of features
    • Broken PDMS and/or glass
    • Uneven PDMS thickness

    All the parameters are controlled under 4X digital microscope and any abnormalities are reported as photo recordings.

    Equipment

    ValueLabel
    Digital MicroscopeNAME
    AndonstarBRAND
    4X ADSM201SKU
    www.nehirbt.com.trLINK

    Microchannel Quality Control

    2.

    The microchannels inside PDMS chips after bonding are needed to be checked for the parameters which are

    • Defects and scratches, micron size
    • Strains and voids, micron size
    • Dirts inside microchannels
    • Airbubbles inside microchannels
    • Multiple channel layers, out of on top alignment
    • Not bond microchannels
    • Incorrect sizes of microchanels and microfeatures

    The microchannel quality control is performed using a high quality microscope and CCD camera. The sizes of micron features are measured using software of the CCD camera.

    Equipment

    ValueLabel
    Epi-Fluorescent MicroscopeNAME
    ZeissBRAND
    AxioObserverSKU
    www.nehirbt.com.trLINK

    Leakage Flow Tests

    3.

    The chip setup is formed by the components given as

    • Syringe pump and 1mL syringe
    • Tubings and Fittings suitable for 2mm ports
    • Microfluidic chip to be testes

    The substeps to follow the leakage test are

    3.1.

    Fill the syringes with 1mL of distilled water.

    Distilled water and isopropanol are two most used chemical for all tests performed in microfluidics.

    The distilled water does not have to be sterilized but it is better to filter the liquid sample.

    3.2.

    Connect the proper fittings on syringe to connect tubing and port of the chips. The proper fittings are chosen from

    http://www.ufluidic.com/collections/fittings-and-connectors

    Mostly, single port is enough to fill all channels of the chip but when multiple ports to be used for filling all channels, distilled water mixed with food coloring is used. Food coloring is also beneficial for detection of the possible leakage points.

    Safety information
    Food coloring is unuseful for very small size channels that is smaller thn 25 microns.

    Equipment

    ValueLabel
    Syringe pumpNAME
    ChemyxBRAND
    Fusion 100SKU
    www.nehirbt.com.trLINK
    3.3.

    The flow rate arranged are increased step by step starting from 0,1 uL/min. The increasins step values are x10 and given as below. When the leakage occurs, the flow rate is recorded. Under normal conditions, leakage must occur higher than 2 mL/min, but chip design and channel resistance effects the mininum flow rate of leakage starts.

    • 0,1 uL/min
    • 1,0 uL/min
    • 10 uL/min
    • 100 uL/min
    • 1mL/min (1000 uL/min)